Sains Malaysiana 43(11)(2014): 1725–1728
Elektroretinogram Multifokus (mfERG): Data Piawai Makmal bagi Kanak-kanak
Berusia 6-12 Tahun
(Multifocal Electroretinogram (mfERG): The Laboratory Standard Values or Children
Aged 6-12 Years Old)
NORLAILA MAT DAUD1*, ROKIAH OMAR1, YANTI ROSLI2 & JAMALIAH RAHMAT3
1Program Optometri & Sains Penglihatan, Pusat Pengajian Sains Jagaan Kesihatan,
Fakulti Sains Kesihatan, Universiti Kebangsaan Malaysia (Kampus KL)
Jalan Raja Muda Abd. Aziz, 50300 Kuala Lumpur, Malaysia
2Program Sains Bioperubatan, Pusat Pengajian Sains Diagnostik Gunaan,
Fakulti Sains Kesihatan, Universiti Kebangsaan Malaysia (Kampus KL), Jalan Raja Muda Abd. Aziz
50300 Kuala Lumpur, Malaysia
3Jabatan Oftalmologi, Hospital Kuala Lumpur, Jalan Pahang, 50300 Kuala Lumpur, Malaysia
Received: 11 July 2013/Accepted: 3 March 2014
ABSTRAK
Objektif kajian ini adalah untuk mendapatkan nilai data piawai mfERG bagi Makmal Elektrofisiologi,
Program Optometri & Sains Penglihatan, Fakulti Sains Kesihatan, Universiti Kebangsaan Malaysia (UKM). Pengukuran mfERG yang dilakukan adalah min ketumpatan amplitud (nV/deg²) kedua mata dan perbandingan min antara mata kanan (OD) dan kiri (OS); respon ketumpatan amplitud setiap bulatan yang dikaji dan respon keseluruhan ketumpatan amplitud kawasan retina yang dikaji. Kajian ini melibatkan seramai 16 orang kanak-kanak normal berumur antara 6 hingga 12 tahun. Ujian t-berpasangan mendapati tiada perbezaan signifikan min ketumpatan amplitud antara kedua mata (OD=24.27±8.16 nV/deg², OS=24.12±8.57 nV/deg², p=0.937). Nilai min ketumpatan amplitud setiap bulatan untuk OD adalah bulatan 1=56.41±24.85 nV/deg², bulatan 2=28.36±9.31 nV/deg², bulatan 3=17.50±5.33 nV/deg², bulatan 4=10.51±3.17 nV/deg² dan bulatan 5=8.57±2.74 nV/deg². Nilai min ketumpatan amplitud setiap bulatan untuk OS adalah bulatan 1=54.67±25.10 nV/deg², bulatan 2=28.89±10.52 nV/deg², bulatan 3=18.23±5.98 nV/deg², bulatan 4=10.40±3.37 nV/deg² dan bulatan 5=8.30±2.73 nV/deg². Manakala min ketumpatan amplitud keseluruhan kawasan retina ialah OD=24.27±8.16 nV/deg² dan OS=24.12±8.57 nV/deg². Kesimpulannya nilai piawai makmal mfERG telah berjaya dihasilkan untuk usia 6 hingga 12 tahun di Makmal Elektroretinogram Multifokal,
Program Optometri dan Sains Penglihatan, UKM. Diharapkan ia akan diguna pakai sebagai panduan dan perbandingan penyelidikan akan datang.
Kata kunci: Elektroretinogram multifokus (mfERG); ketumpatan amplitud; nilai piawai
ABSTRACT
The aim of this study was to get the standard values of multifocal electroretinogram (mfERG)
for the Electrophysiology Laboratory, Optometry Program, Faculty of Health
Sciences, Universiti Kebangsaan Malaysia (UKM). The measurements involved were
mean amplitude density of mfERG in both eyes and to
compare between right and left eye; amplitude density response in each
concentric rings and average of amplitude density of the retina. Sixteen normal
subjects aged between 6-12 years old were involved in this study. Students t-
test showed no significant difference in amplitude density between right and
left eyes (OD=24.27±8.16 nV/deg², OS=24.12±8.57 nV/deg², p=0.937). Mean amplitude
density for each concentric rings (R) in right eye (OD) were R1=56.41±24.85 nV/deg², R2=28.36±9.31 nV/deg²,
R3=17.50±5.33 nV/deg², R4=10.51±3.17 nV/deg² and R5=8.57±2.74 nV/deg².
Mean amplitude density for each concentric rings (R) in left eye (OS)
were R1=54.67±25.10 nV/deg², R2=28.89±10.52 nVV/deg², R3=18.23±5.98 nV/deg²,
R4=10.40±3.37 nV/deg² and R5=8.30±2.73 nV/deg². Mean amplitude density mfERG of the retina were OD = 24.27±8.16 nV/deg²
and OS = 24.12±8.57 nV/deg². In
conclusion, standard values for Multifocal Electroretinogram Laboratory, Program of Optometry & Visual Science, UKM has been developed
for 6-12 years old. We hope these values will serve as a guidance and reference
for future research.
Keywords: Amplitude density; multifocal electroretinogram; standard values
REFERENCES
Fulton, A.B., Hansen, R.M., Moskowitz, A. & Barnaby,
A.M. 2005. Multifocal ERG in subjects with a history
of retinopathy of prematurity. Documenta Ophthalmologica111: 7-13.
Gonzalez,
P., Parks, S., Dolan, F. & Keating, D. 2004. The effects
of pupil size on the multifocal electroretinogram. Doc. Ophthalmol. 109: 67-72.
Hood,
D.C., Bach, M., Brigel, M., Keating, D., Kondo, M.,
Lyons, J.S., Marmor, M.F., McCulloch, D.L. & Palmowski-Wolfe, A.M. 2012. ISCEV
standard for clinical multifocal electroretinography (mfERG) (2011 edition). Doc. Ophthalmol. 124: 1-13.
Hood, D.C., Odel, J.G., Chen, C.S.
& Winn, B.J. 2003. The multifocal electroretinogram. J. Neuro-Ophthamol. 23(3): 225-235.
Hood, D.C., Wladis, E.J., Shady,
S., Holopigian, K., Li, J. & Seiple,
W. 1998. Multifocal rod electroretinogram. Invest. Ophthal. Vis. Sci. 39: 1152-1162.
Kondo, M., Miyake, Y., Horiguchi,
M., Suzuki, S. & Tanikawa, A. 1995. Clinical evaluation of multifocal electroretinogram. Invest. Ophthalmol. Vis. Sci. 36: 2146-2150.
Lai, Y.Y., Chan, W.M., Lai, Y.K., Ngai, W.S., Li, H. &
Lam, S.C. 2007. The clinical applications of
multifocal electroretinography: A systematic review. Survey
of Ophthalmology 52(1): 61-96.
Luu, C.D., Foulds,
W.S. & Tan, T.H. 2007. Features of the multifocal electroretinogram may predict the rate of myopia progression in children. Ophthalmology 114(8):
1433-1438.
Mohidin, N., Yap, K.K.H. & Jacobs, R.J.
1999. Influence of age on the multifocal electroretinography. Ophthal. Physiol.Opt. 19(6): 481-488.
Poloschek, C.M. & Bach, M. 2009. Can we do without mydriasis in multifocal ERG
recordings?. Doc. Ophthalmol. 118: 121-127.
Sutter,
E. & Tran, D. 1992. The field topography of ERG component
in man. Part 1. The photopic luminance response. Vison Res. 32: 433-446.
*Corresponding
author; email: laila6842@yahoo.com.my
|